Correction of spherical aberration in multi-focal multiphoton microscopy with spatial light modulator

Abstract : We demonstrate that high-quality images of the deep regions of a thick sample can be obtained from its surface by multi-focal multiphoton microscopy (MMM). The MMM system incorporates a spatial light modulator to separate the excitation beam into a multi-focal excitation beam and modulate the pre-distortion wavefront to correct spherical aberration (SA) caused by a refractive index mismatch between the immersion medium and the biological sample. When fluorescent beads in transparent epoxy resin were observed using four SA-corrected focal beams, the fluorescence signal of the obtained images was ~52 times higher than that obtained without SA correction until a depth of ~1100 μm, similar to the result for single-focal multiphoton microscopy (SMM). The MMM scanning time was four times less than that for SMM, and MMM showed an improved fluorescence intensity and depth resolution for an image of blood vessels in the brain of a mouse stained with a fluorescent dye.
Type de document :
Article dans une revue
Liste complète des métadonnées

https://hal.univ-lille.fr/hal-03174752
Contributeur : Lilloa Université de Lille <>
Soumis le : vendredi 19 mars 2021 - 14:46:16
Dernière modification le : vendredi 9 juillet 2021 - 11:46:05

Lien texte intégral

Identifiants

Collections

Citation

Naoya Matsumoto, Alu Konno, Yasushi Ohbayashi, Takashi Inoue, Akiyuki Matsumoto, et al.. Correction of spherical aberration in multi-focal multiphoton microscopy with spatial light modulator. Optics Express, Optical Society of America - OSA Publishing, 2017, Optics Express, 25 (6), pp.7055-7068. ⟨10.1364/OE.25.007055⟩. ⟨hal-03174752⟩

Partager

Métriques

Consultations de la notice

20