Effectiveness of in-line filters to completely remove particulate contamination during a pediatric multidrug infusion protocol

Résumé

The large number of drugs administered simultaneously to neonates and children in hospital results in the formation of particles that are potentially infused. We have investigated the ability of IV in-line filters to eliminate particulate matter from multidrug infusion lines and so prevent contamination. The impact on particle occurrence of the internal volume of the IV line below the in-line filter was then evaluated. The multidrug therapy given to children was reproduced with and without in-line filtration. Three combinations with a filter were tested to vary the internal volume (V) between the filter and the catheter egress. The catheter was then connected to a dynamic particle count to evaluate the particulate matter potentially administered to children during infusion. The introduction of in-line filters led to a significant reduction in overall particulate matter, from 416,974 [208,479-880,229] to 7,551 [1,985-11,287] particles (p < 0.001). Larger particles of ≥10 and 25 µm were also significantly reduced. Adding an extension set to the egress of the in-line filter (V = 1.7 mL) caused a significant increase in particulate contamination for both. This study showed that in-line filtration is an effective tool in preventing particle administration to patients. Their position in the infusion in-line is therefore important because of its impact on internal volume and drug particle formation.
Fichier principal
Vignette du fichier
s41598-018-25602-6.pdf (1.5 Mo) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)

Dates et versions

hal-03265646 , version 1 (21-06-2021)

Licence

Paternité

Identifiants

Citer

Maxime Perez, Bertrand Decaudin, Wadih Abou Chahla, Brigitte Nelken, Laurent Storme, et al.. Effectiveness of in-line filters to completely remove particulate contamination during a pediatric multidrug infusion protocol. Scientific Reports, 2018, 8, pp.7714. ⟨10.1038/s41598-018-25602-6⟩. ⟨hal-03265646⟩

Collections

UNIV-LILLE
33 Consultations
30 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More